A pneumatic sampling switch which can be adapted to the outlet of a thermog
ravimetric analyzer furnace with a horizontal purge gas flow was developed.
This system allows a repetitive introduction of shea sampling pulses of th
ermogravimetric analyzer evolved gases to a quadrupole mass spectrometric d
etector. System performance was evaluated through measuring the level of re
peatability under different operation conditions. In addition, pulsed sampl
ing thermogravimetry/mass spectrometry introduced no measurable quantitativ
e or qualitative errors. This system can be used for monitoring thermooxida
tive decomposition reactions, without accelerating the deterioration of qua
drupole mass spectrometric detector performance. Furthermore, it can be use
d as an inlet system for a future on-line TG/GC/MS system. (C) 1999 Elsevie
r Science B.V. All rights reserved.