Plutonium(III)-ion selective electrode of liquid membrane type using multidentate phosphine oxide ionophore

Citation
Y. Kitatsuji et al., Plutonium(III)-ion selective electrode of liquid membrane type using multidentate phosphine oxide ionophore, ANALYT CHIM, 387(2), 1999, pp. 181-187
Citations number
17
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
387
Issue
2
Year of publication
1999
Pages
181 - 187
Database
ISI
SICI code
0003-2670(19990422)387:2<181:PSEOLM>2.0.ZU;2-3
Abstract
Plutonium(III)-ion selective electro de consisting of phosphine oxides such as bis-(diphenylphosphoryl) -methane as an ionophore was developed. Fundam ental data, which were utilized for designing the electrode, on the ion tra nsfer of Pu3+ across aqueous (w)/organic (org) solutions interface facilita ted by the phosphine oxide derivatives were obtained by ion-transfer polaro graphy at the w/og interface. The Pu(III)-ISE prepared exhibited Nernstian response to the concentration of Pu3+ in the sample solution. Interference of UO22+, U4+, NpO2+ and Pu4+ was investigated. (C) 1999 Elsevier Science B .V. All rights reserved.