M. Baleani et al., Endurance testing of hip prostheses: a comparison between the load fixed in ISO 7206 standard and the physiological loads, CLIN BIOMEC, 14(5), 1999, pp. 339-345
Background. pre-clinical endurance validation of innovative hip prostheses
femoral components are carried out following the ISO 7206 standard. The in
vitro fatigue test must simulate accurately the physiological load to corre
ctly define the section of the stem.
Objective. This study investigates if the loads defined in the ISO 7206 sta
ndard simulate correctly the physiological load that occur in vivo.
Design. Simulation of the damage induced by in vitro test and in vivo loads
is performed. Different designs of the stem are examined. Materials with d
ifferent fatigue limit are considered. For the in vivo loads, different bod
y weights are modelled.
Methods. The minimal stem dimensions required to stand 20 years of patient
use and to pass the fatigue test are calculated based on linear damage accu
mulation.
Results. The results show that the ISO load simulates reasonably well the p
hysiological load that is likely to occur in active patients with a low or
normal body weight. Conversely, the ISO test underestimates the physiologic
al load that heavy and active patients are likely to apply to the implant.
Conclusions. Different minimum requirements for endurance strength should b
e fixed for hip prostheses designed for patients of different weight. This
solution would reduce the risk of undersizing or oversizing the stem.