Objective. To evaluate the accuracy in determining centre of pressure when
using a force platform for stabilometry research.
Design. Technical report.
Background. Previous studies have determined the accuracy of locating the c
entre of pressure by applying force through a single point, In human stabil
ometry research however, force is distributed across two feet and the centr
e of pressure located between the feet.
Methods. Forces were applied to a Kistler force platform using a force appl
ied through a single point as well as when force was distributed between tw
o metal blocks.
Results. Errors in calculating centre of pressure were substantially larger
when force was applied through a single point than when the force was dist
ributed between two blocks.
Conclusions. Evaluating centre of pressure accuracy by applying force throu
gh a single point overestimates the errors inherent in human stabilometry r
esearch.