A simple method for the simultaneous determination of five artificial sweet
eners, alitame (AL), acesulfame K (AK), saccharin (SA), aspartame (APM) and
dulcin (DU) in various foods by high performance liquid chromatography (HP
LC) was developed.
Chopped or homogenized samples were packed into cellulose tubing with 0.01
mol/L hydrochloric acid containing 10% sodium chloride, and dialyzed agains
t 0.01 mol/L hydrochloric acid for 24 similar to 48 hours. Tetra-n-butylamm
onium bromide and pH 5.0 phosphate buffer were added to the dialyzate. The
solution was passed through a Sep-Pak Vac C18 cartridge, and the cartridge
was washed with water and a mixture of methanol-water (1:9). The five sweet
eners were eluted from the cartridge with a mixture of methanol-water (45:5
5). The sweeteners were separated on an Inertsil ODS-2 column with a mobile
phase of methanol-water (1:3) containing 0.01 mol/L tetra-n-propylammonium
hydroxide adjusted to pH 3.5 with phosphoric acid and were detected at 210
nm.
The recoveries of the five sweeteners from various kinds of foods spiked at
200 mu g/g ranged from 77 similar to 102%. The detection limits of the fiv
e sweeteners were 10 mu g/g in the samples.