Multicenter evaluation of electrical stimulation systems for walking

Citation
M. Wieler et al., Multicenter evaluation of electrical stimulation systems for walking, ARCH PHYS M, 80(5), 1999, pp. 495-500
Citations number
21
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine
Journal title
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION
ISSN journal
00039993 → ACNP
Volume
80
Issue
5
Year of publication
1999
Pages
495 - 500
Database
ISI
SICI code
0003-9993(199905)80:5<495:MEOESS>2.0.ZU;2-B
Abstract
Objective: To test the long-term benefits of several noninvasive systems fo r functional electrical stimulation (FES) during walking. Design: Forty subjects (average years since injury, 5.4) were studied in fo ur centers for an average time of 1 year. Gait parameters were tested for a ll subjects with and without FES. Thus, subjects served as their own contro ls, since the specific effect of using FES could be separated from improvem ents resulting from other factors (eg, training). Setting: Subjects used the devices in the community, but were tested in a u niversity or hospital setting. Patients: Subjects with spinal cord injury (n = 31) were compared to subjec ts with cerebral damage (n = 9). Main Outcome Measures: Gait parameters (speed, cycle time, stride length). Acceptance was studied by means of a questionnaire. Results: Some initial improvement in walking speed (average increase of >20 %) occurred, and continuing gains were seen (average total improvement, 45% ). The largest relative gains were seen in the slowest walkers (speeds of < 0.3misec). Acceptance of the FES systems was good and improved systems have been developed using feedback from the subjects. Conclusions: Based on the improvements in speed and the acceptance of these FES systems, a greatly increased role for FES in treating gait disorders i s suggested. (C) 1999 by the American Congress of Rehabilitation Medicine a nd the American Academy of Physical Medicine and Rehabilitation.