There are many potential applications of three-dimensional (3D) image regis
tration in MR-guided radiofrequency (RF) thermal ablation of tumors. For ex
ample, after registration of image volumes obtained before and after therma
l ablation, a variety of quantitative and visual assessments of therapy wer
e performed, For liver tumors, the accuracy of rigid body, manual registrat
ion on 19 pairs of image volumes was evaluated, almost all of which were ob
tained during thermal treatment sessions. Registration error was estimated
as a distance between anatomical landmarks, including both internal vascula
r structures as well as the surface of the Liver. Over all image pairs and
throughout a large portion of the liver, the registration error was 3.1 mm
(mean + I SD), From the bottom to the top of the liver, error increased on
the order of 13%, probably because of liver motion and deformation resultin
g from respiration, Although probably insufficient for blind guidance of th
erapy, registration accuracy was undoubtedly sufficient for interesting app
lications in the planning, assessment, and optimization of interventional M
R-guided thermal treatment of liver tumors.