Ma. Proskurnin et Ag. Abroskin, Optimization of optical system parameters in dual-beam thermal lens spectrometry, J ANALYT CH, 54(5), 1999, pp. 401-408
Citations number
16
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
10619348 → ACNP
Volume
54
Issue
5
Year of publication
1999
Pages
401 - 408
Database
ISI
SICI code
1061-9348(199905)54:5<401:OOOSPI>2.0.ZU;2-L
Abstract
An equation is derived for the interrelationship between the thermal-lens s ignal and geometry parameters of the optical system of the dual-beam spectr ometer (distances between beam waists, sample cell, and the detector). The optimization of the optical system allowed the limits of detection and dete rmination of Methyl Orange as an example model system to be lowered by an o rder of magnitude and the linear calibration range to be widened by more th an one order of magnitude.