Results obtained by the Diagnostic Systems Laboratories (DSL), I-125 direct
androstenedione kit performed according to the package insert and after th
e introduction of an extraction step were compared with a reference method
that used a tritiated radioligand, Significantly higher results were obtain
ed by the direct assay as compared to the tritiated method, whereas excelle
nt correlation was obtained between the two after the introduction of an ex
traction step in the direct procedure. The discordance between the results
by the DSL direct and extraction methods may be due to matrix effects and n
on-specific interferences, The performance characteristics of the kit were
evaluated and were found to be satisfactory. Twenty apparently healthy indi
viduals gave a range of 26-228 ng/dL for males and 36-162 ng/dL for females
in the DSL extraction assay.