Improving resolution in capillary zone electrophoresis through bulk flow control

Citation
S. Kar et Pk. Dasgupta, Improving resolution in capillary zone electrophoresis through bulk flow control, MICROCHEM J, 62(1), 1999, pp. 128-137
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
MICROCHEMICAL JOURNAL
ISSN journal
0026265X → ACNP
Volume
62
Issue
1
Year of publication
1999
Pages
128 - 137
Database
ISI
SICI code
0026-265X(199905)62:1<128:IRICZE>2.0.ZU;2-A
Abstract
One can increase the residence time of analytes in a capillary electrophore sis experiment by reducing the electroosmotic flow. Such measures increase separation efficiency and resolution. It is posible to accomplish this by h ydrostatic backflow but dispersion induced by the parabolic flow profile ca uses considerable plate losses except with extremely small diameter capilla ries. Relative to hydrostatic backflow, auxiliary electroosmotic pumping, d eployed to increase residence time, produces better resolution at the same residence time and thus constitutes a superior alternative, (C) 1999 Academ ic Press.