E. Filippi et al., Hourly measurement of particulate concentrations with streaker samplers and optical methods, NUCL INST B, 150(1-4), 1999, pp. 370-374
We have studied the possibility of continuous particulate monitoring, by an
optical system, on the filters of a two-stage streaker sampler. The streak
er deposits, collected for PIXE analysis, have been examined off-line by me
asuring the attenuation of a light beam. The optical analysis provides, in
a few minutes, an estimate of hourly total mass concentrations, over a one
week period, for both stages, very useful as a selection for further PIXE i
rradiation. We describe the set-up and the calibration procedures, where PI
XE plays a fundamental role in resolving different aerosol compositions and
in validating the prototype performance, The results obtained over 2 weeks
of operation and the comparison with the aerosol component determined via
PIXE analysis, are also discussed. (C) 1999 Elsevier Science B.V. All right
s reserved.