Analysis of the best site on the stapes footplate for ossicular chain reconstruction

Citation
M. Asai et al., Analysis of the best site on the stapes footplate for ossicular chain reconstruction, ACT OTO-LAR, 119(3), 1999, pp. 356-361
Citations number
11
Categorie Soggetti
Otolaryngology,"da verificare
Journal title
ACTA OTO-LARYNGOLOGICA
ISSN journal
00016489 → ACNP
Volume
119
Issue
3
Year of publication
1999
Pages
356 - 361
Database
ISI
SICI code
0001-6489(1999)119:3<356:AOTBSO>2.0.ZU;2-E
Abstract
Experiments were performed in 22 fresh human temporal bones to compare the relative acoustic function of three stapes footplate sites for an incus sta pes superstructure replacement prosthesis (I-SRP). The three sites evaluate d were the anterior, centre and posterior footplates. A new round window (R W) measurement method was used to make the comparisons. A small glass micro sphere was placed in the centre of the RW as a target. A Polytec laser Dopp ler vibrometer was used to measure round window displacement in response to 50 pure tones between 200 and 10,000 Hz presented at 80 dB SPL at the tymp anic membrane (TM). After a baseline measurement of RW displacement in the intact temporal bone, the incus was removed and a cement I-SRP (CIRP) forme d between the mid-malleus handle and each of the three test footplate sites , in random order. RW displacement was again measured after placement of th e CIRP at each of three sites. We found the centre site to be 3.0-7.0 dB be tter than the anterior site above 2,000 Hz. There were no differences betwe en the anterior and centre sites below 2,000 Hz. The posterior site was the worst at all frequencies.