Experiments were performed in 22 fresh human temporal bones to compare the
relative acoustic function of three stapes footplate sites for an incus sta
pes superstructure replacement prosthesis (I-SRP). The three sites evaluate
d were the anterior, centre and posterior footplates. A new round window (R
W) measurement method was used to make the comparisons. A small glass micro
sphere was placed in the centre of the RW as a target. A Polytec laser Dopp
ler vibrometer was used to measure round window displacement in response to
50 pure tones between 200 and 10,000 Hz presented at 80 dB SPL at the tymp
anic membrane (TM). After a baseline measurement of RW displacement in the
intact temporal bone, the incus was removed and a cement I-SRP (CIRP) forme
d between the mid-malleus handle and each of the three test footplate sites
, in random order. RW displacement was again measured after placement of th
e CIRP at each of three sites. We found the centre site to be 3.0-7.0 dB be
tter than the anterior site above 2,000 Hz. There were no differences betwe
en the anterior and centre sites below 2,000 Hz. The posterior site was the
worst at all frequencies.