Mm. Firozabadi et Vi. Ushakov, One-electron mode of Phi(sic)y-130 multiplier tubes and detection of Cerenkov radiation photons, INSTR EXP R, 42(2), 1999, pp. 192-199
Methods for the detection of single photons with photomultiplier tubes (PMT
s), semiconductor, and combined detectors are compared. The one-electron ch
aracteristics of modern detectors and the analytical relationships between
the one- and multi-electron amplitude resolution and the dynode secondary-e
mission coefficients are presented. The pulse height spectra of short light
pulses (tau = 1.5-15 ns) produced by a setup with an avalanche generator w
ere measured with Phi(sic)Y-130 PMTs operating in the one-electron mode, an
d this technique was shown to be promising for the detection of Cerenkov ra
diation. With the aim of using Phi sicY-130 PMTs in differential Cerenkov c
ounters, methods to further improve their one-electron characteristics were
considered.