T. Hankemeier et al., Optimization of large-volume on-column injection conditions in gas chromatography by monitoring the actual carrier gas flow, J CHROMAT A, 841(1), 1999, pp. 75-94
The change of the evaporation rate of the solvent during injection and evap
oration is the most critical aspect during optimization of large-volume on-
column injection conditions in gas chromatography. The change is caused by
the pressure drop along the retention gap when using an early solvent vapou
r exit (SVE) and can be described by a mathematical model. Four procedures
for the optimization of the injection conditions were compared. It was foun
d that different procedures often yield different evaporation rates, which
may also depend on the injection speeds used during optimization. For optim
ization of a new set-up, i.e. if little is known about the optimal injectio
n conditions, the evaporation rate should be determined by increasing the i
njection time at a fixed injection speed, injection temperature and head pr
essure; subsequently, an appropriate injection speed can be calculated. If
a mere re-optimization is required as e.g, after the exchange of the retent
ion gap, adjusting the evaporation rate to the injection speed by varying t
he injection temperature at a constant injection speed is the preferred pro
cedure. With both methods, optimization can be achieved by means of 2-5 inj
ections of pure solvent and monitoring the helium carrier gas flow. That is
, optimization of the injection conditions takes less than 1 h. When using
this strategy, analytes as volatile as monochlorobenzene can be determined
in aqueous samples by in-vial liquid-liquid extraction-gas chromatography-m
ass spectrometry. Closing the SVE at the very end of solvent evaporation re
sults in a considerable increase of the capacity of the retention gap compa
red to closing the SVE before all solvent is evaporated. (C) 1999 Elsevier
Science B.V. All rights reserved.