Ja. Moreton et Ht. Delves, Use of Virkon as a disinfectant for clinical samples carrying a high risk of infection in inductively coupled plasma mass spectrometry, J ANAL ATOM, 14(5), 1999, pp. 893-894
The addition of 1% Virkon, a virucidal disinfectant, to clinical samples ca
rrying a high risk of infection can reduce the likelihood of infection, par
ticularly when samples are nebulised, as in ICP-MS. Semiquantitative analys
is of Virkon by ICP-MS showed the major elements present in Virkon (Na, Ca,
Mg, Al and Zr), which would preclude their determination in samples to whi
ch Virkon had been added, and those elements present at intermediate levels
(Br, B, Zn, Sc, Mo and Cr) where accurate determination by ICP-MS would be
difficult. With suitable blank subtraction, it would be possible to add Vi
rkon to high risk biological samples for the determination of many elements
by ICP-MS and AAS.