Nd. Yordanov et al., Studies on some materials suitable for use as internal standards in high energy EPR dosimetry, J RAD NUCL, 240(2), 1999, pp. 619-622
EPR-spectroscopic properties (line-intensity, -width, and g-factors) of pyr
olized at 550 degrees C sucrose, MgO and MgO doped with Mn2+ ions (500:1) r
emain unchanged after high-dose (1-100 kGy) irradiation, whereas CaO gives
an EPR signal. These properties of these materials make it possible to use
them as internal reference EPR standards in the work under conditions of st
rong ionisation environment, for precise determination of the G-values of o
ther materials, for obtaining exact magnitudes of increase of the EPR respo
nse when the method of additional dose is applied, to follow the kinetics o
f decay of some radiation induced defects.