Performance of n-gamma pulse-shape discrimination with simple pile-up rejection at high gamma-ray count rates

Citation
T. Okuda et al., Performance of n-gamma pulse-shape discrimination with simple pile-up rejection at high gamma-ray count rates, NUCL INST A, 426(2-3), 1999, pp. 497-502
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
426
Issue
2-3
Year of publication
1999
Pages
497 - 502
Database
ISI
SICI code
0168-9002(19990501)426:2-3<497:PONPDW>2.0.ZU;2-I
Abstract
The performance of n-gamma pulse-shape discrimination for a liquid scintill ation detector has been investigated for gamma-ray count rates up to 50 kcp s. A method in which the ratio of the total to partial charge in the anode pulse is directly measured has shown much improved quality of the pulse-sha pe discrimination when pile-up events are rejected; it can discriminate neu tron events of 50 cps from gamma-ray events of 29 kcps. The method with sim ple pile-up rejection has the advantage that only general purpose electroni cs are required. (C) 1999 Elsevier Science B.V. All rights reserved.