On-line monitoring of calcium in natural, borehole and boiler feed water with a flow injection system using a calcium-selective membrane electrode asa sensor
Jf. Van Staden et Ri. Stefan, On-line monitoring of calcium in natural, borehole and boiler feed water with a flow injection system using a calcium-selective membrane electrode asa sensor, S AFR J C-S, 52(1), 1999, pp. 24-26
Citations number
18
Categorie Soggetti
Chemistry
Journal title
SOUTH AFRICAN JOURNAL OF CHEMISTRY-SUID-AFRIKAANSE TYDSKRIF VIR CHEMIE
An on-line automated system for the determination of calcium in natural, bo
rehole and boiler feed water based on the concept of flow injection with a
calcium-selective membrane electrode as sensor is described. The system is
suitable for the on-site monitoring of calcium at a sampling rate of 60 sam
ples per hour in the linear range 10(-4)-10(-2) mol l(-1)(approximately 4-4
00 mg l(-1)) with an RSD better than 0.7%. The detection limit of the syste
m is 2.1 x 10(-6) mol l(-1) (about 0.084 mg l(-1)).