H. Du Plessis et al., Simultaneous determination of certain selected trace elements with an on-line preconcentrating flow injection ICP-OES system, S AFR J C-S, 52(1), 1999, pp. 27-35
Citations number
16
Categorie Soggetti
Chemistry
Journal title
SOUTH AFRICAN JOURNAL OF CHEMISTRY-SUID-AFRIKAANSE TYDSKRIF VIR CHEMIE
The detection limits of an ICP instrument were improved by on-line preconce
ntrating samples on a miniature ion exchange column incorporated into the c
onduits of a flow injection (FI) system. Several operating parameters for b
oth the ICP and the ion exchange column were optimised to maximise the adva
ntages of the preconcentrating system. These include the pH and loading rat
e of the sample, the length and diameter of the column and the eluent conce
ntration. The preconcentrating FIA-ICP system gave simultaneous multi-eleme
nt detection limits, which were between five and 30 times better than for c
onventional continuously aspirating systems for the elements barium, calciu
m, chromium, copper, manganese, iron and lead. Precision was better than 3.
5% RSD at the 50 mu g l(-1) level for Ba, Ca, Cr, Cu, Fe and Mn, and the 25
0 mu g l(-1) level for Pb. Analysis of spiked rain water in the mu g l(-1)
level gave nearly 100% recovery. The sampling rate was six samples per hour
.