Analysis of high purity antimony by glow discharge quadrupole mass spectrometry

Citation
R. Shekhar et al., Analysis of high purity antimony by glow discharge quadrupole mass spectrometry, ATOM SPECT, 20(1), 1999, pp. 25-29
Citations number
8
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ATOMIC SPECTROSCOPY
ISSN journal
01955373 → ACNP
Volume
20
Issue
1
Year of publication
1999
Pages
25 - 29
Database
ISI
SICI code
0195-5373(199901/02)20:1<25:AOHPAB>2.0.ZU;2-Q
Abstract
Characterization of a high-purity antimony sample for its trace and ultratr ace level impurities has been carried out using GD-QMS. Relative sensitivit y factor (RSF) values generated from the analysis of NIST standard referenc e materials (Low Alloy Steel SRM 1761 and SRM 1762) were used to compute th e concentrations. These RSF values were validated through the analysis of N IST Low Alloy Steel SRM 1766. The impurity levels determined through GD-QMS were cross-validated with other solution-based techniques, namely ICP-QMS, ICP-OES, and GFAAS. Good agreement was observed for many elements where in terferences were not noted.