On calibration of solid phase microextraction-gas chromatography-mass spectrometry system for analysis of organic air contaminants using gaseous standard mixtures
J. Namiesnik et al., On calibration of solid phase microextraction-gas chromatography-mass spectrometry system for analysis of organic air contaminants using gaseous standard mixtures, CHEM ANAL, 44(2), 1999, pp. 201-213
The paper briefly touches the general problems related to calibration in an
alytical measurements. The special requirements with respect to calibration
in methods of air organic contaminants determination (those based on solid
phase microextraction, SPME) are stressed. A convenient apparatus for gene
ration of gaseous mixture with low organic analyte content with facilities
for SPME sampling is described. The procedure for temperature and humidity
corrected calibration is proposed and studies on effects of these parameter
s on calibration curves are presented.