Improvement of the detection limit in capillary gas chromatographic trace analysis of C-2-C-4 hydrocarbons by post column cryogenic trapping

Citation
M. Habram et T. Welsch, Improvement of the detection limit in capillary gas chromatographic trace analysis of C-2-C-4 hydrocarbons by post column cryogenic trapping, HRC-J HIGH, 22(6), 1999, pp. 335-338
Citations number
15
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY
ISSN journal
09356304 → ACNP
Volume
22
Issue
6
Year of publication
1999
Pages
335 - 338
Database
ISI
SICI code
0935-6304(199906)22:6<335:IOTDLI>2.0.ZU;2-J
Abstract
A method for the improvement of signal-to-noise ratios and hence detection limits in capillary gas chromatographic trace analysis of C-2-C-4 hydrocarb ons was developed. It is based on the post column cryogenic trapping of the separated hydrocarbons using liquid nitrogen as coolant and its fast reinj ection into the detector by rapid heating of the capillary tubing used for trapping. The improved signal-to-noise ratios were applied to decrease the sample volume and/or to lower the detection limit 10 to 27 times. The conce ntrations of these hydrocarbons in an urban air sample determined without a nd with post column cryogenic trapping were in good agreement but the preci sion was better when applying the cryogenic trapping mode.