Software error resulting in malfunction of an implantable cardioverter defibrillator

Citation
Ma. Coppess et al., Software error resulting in malfunction of an implantable cardioverter defibrillator, J CARD ELEC, 10(6), 1999, pp. 871-873
Citations number
3
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY
ISSN journal
10453873 → ACNP
Volume
10
Issue
6
Year of publication
1999
Pages
871 - 873
Database
ISI
SICI code
1045-3873(199906)10:6<871:SERIMO>2.0.ZU;2-8
Abstract
Software Error. Loss of ventricular output resulting from an unexpected sof tware error in a dual chamber implantable cardioverter defibrillator (ICD) is reported, A 70-year-old man with a dual chamber ICD implanted for a hist ory of cardiac arrest and infra-Hisian bloch-presented with acute onset of dizziness. He was found to have loss of ventricular output due to an intern al software problem, The problem was corrected by software reprogramming vi a the programmer. This malfunction exemplifies the potential ability to cor rect current-generation ICD software problems noninvasively, thus avoiding the need for replacement.