Evaluation of column performance in constant pressure and constant flow capillary gas chromatography

Citation
Lm. Blumberg et al., Evaluation of column performance in constant pressure and constant flow capillary gas chromatography, J CHROMAT A, 842(1-2), 1999, pp. 15-28
Citations number
32
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
842
Issue
1-2
Year of publication
1999
Pages
15 - 28
Database
ISI
SICI code
Abstract
It has been demonstrated that there is no significant difference between th e maximum peak capacity and analysis time at the same peak capacity in cons tant pressure (isobaric) and constant flow (isorheic) modes of temperature programmed GC. In view of that, the choice between the two pneumatic modes should be based on considerations other than the speed-separation performan ce of the column. (C) 1999 Elsevier Science B.V. All rights reserved.