MEMORY ARTIFACT RELATED TO SELENIUM-BASED DIGITAL RADIOGRAPHY SYSTEMS

Citation
Hg. Chotas et al., MEMORY ARTIFACT RELATED TO SELENIUM-BASED DIGITAL RADIOGRAPHY SYSTEMS, Radiology, 203(3), 1997, pp. 881-883
Citations number
16
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00338419
Volume
203
Issue
3
Year of publication
1997
Pages
881 - 883
Database
ISI
SICI code
0033-8419(1997)203:3<881:MARTSD>2.0.ZU;2-K
Abstract
Digital images acquired on radiography systems with amorphous selenium detectors are susceptible to ''memory artifacts'' from prior x-ray ex posures. In routine clinical use and in a laboratory experiment, artif acts appeared in chest radiographs until the selenium recovered from i nitial exposure. Memory artifacts were eliminated when 3 minutes or mo re elapsed between acquisition of a lateral chest radiograph and acqui sition of the next radiograph.