Metallized sheathless electrospray emitters for use in capillary electrophoresis orthogonal time-of-flight mass spectrometry

Citation
Dr. Barnidge et al., Metallized sheathless electrospray emitters for use in capillary electrophoresis orthogonal time-of-flight mass spectrometry, RAP C MASS, 13(11), 1999, pp. 994-1002
Citations number
16
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
ISSN journal
09514198 → ACNP
Volume
13
Issue
11
Year of publication
1999
Pages
994 - 1002
Database
ISI
SICI code
0951-4198(1999)13:11<994:MSEEFU>2.0.ZU;2-1
Abstract
Chromiun-gold coated fused silica emitters for electrospray ionization have been fabricated with lifetimes upwards of 100 hours. The coating on these emitters is extremely durable withstanding electrical discharges, abrasion and mechanical stress, Usable flow rates range from 30 nL/min to 1 mu L/min using both standard electrospray solutions and aqueous buffers. Coated cap illary electrophoresis columns incorporating the emitter provide excellent separation, low limits of detection (70-700 attomoles fur peptides), and ar e easy to use. performance of the emitters in continuous infusion experimen ts and capillary electrophoresis/ electrospray ionization mass spectrometry applications are presented. Copyright (C) 1999 John Wiley & Sons, Ltd.