Experimental studies on a compact electron cyclotron resonance plasma x-ray source

Citation
R. Baskaran et Ts. Selvakumaran, Experimental studies on a compact electron cyclotron resonance plasma x-ray source, REV SCI INS, 70(6), 1999, pp. 2637-2645
Citations number
31
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
70
Issue
6
Year of publication
1999
Pages
2637 - 2645
Database
ISI
SICI code
0034-6748(199906)70:6<2637:ESOACE>2.0.ZU;2-F
Abstract
A compact electron cyclotron resonance plasma x-ray source, which has a pot ential use for medical imaging is presented in this article. In this articl e, the experimental system and the characterization studies on plasma and x ray are presented. Using a Langmuir probe, the plasma parameters are measu red for different magnetic field profiles and gas pressures. The x-ray spec trum is obtained for various gas pressures and magnetic field profiles. In the x-ray spectrum, the Bremstrahlung radiation, peaking at 20-60 keV is ob served and the final energy of the x ray is extended up to similar to 200 k eV. Thermo luminescence dosimeter (CaSO4 sample) is used for estimating the dose at the port and these results are presented for typical x-ray spectra . Using a teletector, the dose at the port for various coil current is meas ured and these are compared with the estimated dose obtained from the x-ray spectrum. (C) 1999 American Institute of Physics. [S0034-6748(99)05006-6].