Characteristics of 63.6 degrees cylindrical energy analyzer used as chargeexchange neutral particle analyzer

Citation
T. Goto et al., Characteristics of 63.6 degrees cylindrical energy analyzer used as chargeexchange neutral particle analyzer, REV SCI INS, 70(6), 1999, pp. 2661-2664
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
70
Issue
6
Year of publication
1999
Pages
2661 - 2664
Database
ISI
SICI code
0034-6748(199906)70:6<2661:CO6DCE>2.0.ZU;2-Z
Abstract
A charge exchange neutral particle analyzer including a newly designed mult ichannel 63.6 degrees cylindrical electrostatic energy analyzer was constru cted. Since the energy analyzer can simultaneously adopt a small radius of cylindrical electrodes and a large width of an entrance slit, high incident current can be analyzed while the size of the analyzer remain compact. Cha racteristics of the analyzer were investigated using hydrogen beam with ene rgies from about 0.5 to 5 keV. The experimental results are compared well w ith the numerical analysis. (C) 1999 American Institute of Physics. [S0034- 6748(99)01906-1].