Improved calibrator for the evaluation of batch produced strain gauges in cryogenic environments

Citation
Wh. Huber et al., Improved calibrator for the evaluation of batch produced strain gauges in cryogenic environments, REV SCI INS, 70(6), 1999, pp. 2872-2874
Citations number
9
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
70
Issue
6
Year of publication
1999
Pages
2872 - 2874
Database
ISI
SICI code
0034-6748(199906)70:6<2872:ICFTEO>2.0.ZU;2-7
Abstract
We have constructed a simple, improved apparatus for calibrating strain gau ges down to liquid helium temperatures to permit an estimate of the repeata bility and linearity of commercially available, batch produced, metal foil strain gauges. The apparatus is based upon an original design first propose d by McClintock [R. M. McClintock, Rev. Sci. Instrum. 30, 715 (1959)]. Cali brations performed on a particular batch produced metal foil strain gauge i ndicate that it behaved ideally at low temperatures. Although the gauge res istance was strongly temperature dependent, the gauge factor was virtually independent of temperature and did not display any detectable nonlinearity or hysteresis. (C) 1999 American Institute of Physics. [S0034-6748(99)04506 -2].