A miniaturized analysing system for the rapid assay of liver transaminases
activities was set up by means of hybrid technology. Microfluidics was real
ized with printed circuit board technology and combined with thin film glut
amate biosensors. The function oft his system was demonstrated in buffer so
lutions with GOT (AST) and GPT (ALT). The sample and reagent volumes requir
ed are 70 mu l each and the assay time is 4.5 minutes. The miniaturized sys
tem has an outer dimensions of 42 x 22 x 1.5 mm(3) and a total internal vol
ume of 11 mu l.