We carried out 71 primary total hip arthroplasties using porous-coated, hem
ispherical press-fit Duraloc '100 Series' cups in 68 consecutive patients;
61 were combined with the cementless Spotorno stem and ten with the cemente
d Lubinus SP II stem. Under-reaming of 2 mm achieved a press-fit. Of the 71
hips, 69 (97.1%) were followed up after a mean of 2.4 years. Migration ana
lysis was performed by the fin Bild Rontgen Analyse method, with an accurac
y of 1 mm,
The mean total migration after 24 months was 1.13 mm, Using the definition
of loosening as a total migration of 1 mm, it follows that 30 out of 63 cup
s (48%) were loose at 24 months.