Cm. Sunta et al., Anomalies in the determination of the activation energy of thermoluminescence glow peaks by general-order fitting, J PHYS D, 32(11), 1999, pp. 1271-1275
Thermoluminescence glow curves calculated from three physically meaningful
models are fitted into the empirical general-order (GO) kinetics expression
to find the fitted values of activation energy E and the kinetic order b.
The results show that the figure of merit of the fit deteriorates progressi
vely as the best fitted value of b deviates from 1 and 2. At the same time
the fitted E value also departs from its true value. This result is used to
reinterpret the observed change in E of glow peak 5 of LiF (TLD-100) in po
st-irradiation annealed samples. It is concluded that the application of GO
kinetics to glow peak 5 of LiF (TLD-100) leads to inconsistencies and that
the changes in E found by GO fitting may not be real.