Comparison of the NIST and BIPM air-kerma standards for measurements in the low-energy x-ray range

Citation
Dt. Burns et al., Comparison of the NIST and BIPM air-kerma standards for measurements in the low-energy x-ray range, J RES NAT I, 104(2), 1999, pp. 135-140
Citations number
7
Categorie Soggetti
Multidisciplinary,"Engineering Management /General
Journal title
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
ISSN journal
1044677X → ACNP
Volume
104
Issue
2
Year of publication
1999
Pages
135 - 140
Database
ISI
SICI code
1044-677X(199903/04)104:2<135:COTNAB>2.0.ZU;2-I
Abstract
A direct comparison was made between the air-kerma standards used for the m easurement of low-energy x rays at the National Institute of Standards and Technology (NIST) and the Bureau International des Poids et Mesures (BIPM). The comparison was carried out at the BIPM using the BIPM reference beam q ualities in the range from 10 kV to 100 kV. The results show the standards to be in agreement to around 0.5 % at reference beam qualities up to 50 kV and at 100 kV. The result at the 80 kV beam quality is less favorable, with agreement at the 1 % level.