R. Matheoud et al., Changes in the energy response of a dedicated gamma camera after exposure to a high-flux irradiation, PHYS MED BI, 44(6), 1999, pp. N129-N135
This work reports the effects of the gain variation of the photomultiplier
tubes (PMTs) observed on a cardiac dedicated gamma camera after accidental
high-flux irradiation. One detector of this dual-headed 90 degrees-fixed ga
mma camera was accidentally left uncollimated during a quality assurance pr
ocedure on the other detector with a Co-57 flood source (259 MBq) and recei
ved a non-uniform high flux of 1.9-0.6 Mcps over 25 000 mm(2) areas for abo
ut 30 min. To evaluate the severity and the duration of the perturbation ef
fect on the energy response of the detector, the photopeak position was mon
itored for about 1 month with a Tc-99m point source. The 140 keV photopeak
shifted to 158 keV soon after irradiation, reached the correct position aft
er 9 days and moved to a stable value of 132 keV after 15 days. Afterwards,
a new energy calibration reset the photopeak position at 140 keV and the c
orrect energy response of the gamma camera. This experience suggests that p
articular care should be taken to avoid exposures to high radiation fluxes
that induce persistent gain shifts on the PMTs of this system.