J. Pannek et al., An unexpected complication of the remote-controlled intraurethral valve pump for urinary incontinence, UROL INTERN, 61(4), 1998, pp. 235-236
We report an unexpected complication of the remote-controlled intraurethral
valve pump in a patient with neurogenic bladder, A meningomyelocele patien
t with an atonic bladder received an intraurethral valve pump. Acute urinar
y retention was caused by a mucus clot obstructing the pump. Prior to exten
ded clinical use, the remote-controlled intraurethral valve pump should be
evaluated in prospective multicenter studies.