R. Tramontano et Vr. Vanin, Mixing activity-calibrated and uncalibrated gamma-ray sources in efficiency calibration, APPL RAD IS, 51(3), 1999, pp. 323-328
A method to calibrate the absolute efficiency of HPGe detectors in a large
energy range using both activity-calibrated and uncalibrated multienergy ga
mma-ray radioactive sources is presented. The adopted efficiency calibratio
n function epsilon(E) has the simple form In epsilon(E) = a(0) + a(1) In(E/
E-B)+ a(2) In-2(E/E-B). Different parameters a(0) are simultaneously fitted
to the activity-calibrated and uncalibrated sources experimental efficienc
y data by the least-squares method. As an application example, the calibrat
ion of a HPGe detector in the 500 to 3500 keV energy range using a Bi-207 a
ctivity calibrated source and one uncalibrated Co-56 source is described. (
C) 1999 Elsevier Science Ltd. All rights reserved.