A simple and rapid method using liquid chromatography/electrospray ionizati
on-mass spectrometry (LC/ESI-MS) has been developed for the determination d
f bisphenol A in canned drinks. LC separation was carried out on a Zorbax X
DB-C18 column (150 x 2.1 mm) using 0.01% acetic acid-acetonitrile (60 : 40)
as the mobile phase at a flow rate of 0.2 ml/min. The negative ionization
produced a typical (M - H)(-) molecular ion of bisphenol A (m/z 227). The c
alibration graph for bisphenol A was rectilinear from 10 to 1000 pg with a
detection limit by selected ion monitoring (SIM) of about 5 pg. Samples wer
e cleaned up on a Isolute Multimode cartridge (500 mg). The recovery of bis
phenol A from beverages fortified at levels 5 and 50 ng/ml was 85.4 similar
to 96.3%, with high precision. Bisphenol A concentrations of canned drinks
were N.D. similar to 212.1 ng/ml. The limit of quantification of bisphenol
A in canned drinks was 0.5 ng/ml, except for coffee drinks.