Measurement techniques and results of an intercomparison for RF power in a3.5 mm coaxial line up to 26 GHz

Citation
D. Janik et al., Measurement techniques and results of an intercomparison for RF power in a3.5 mm coaxial line up to 26 GHz, IEEE INSTR, 48(2), 1999, pp. 166-168
Citations number
4
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
48
Issue
2
Year of publication
1999
Pages
166 - 168
Database
ISI
SICI code
0018-9456(199904)48:2<166:MTAROA>2.0.ZU;2-I
Abstract
The methods and results of an international comparison are presented for me asurements of RF power in a 3.5 mm coaxial line at seven frequencies up to 26 GHz. Three national laboratories of metrology having different measureme nt techniques participated in the comparison. The agreement between the mea n values of the calibration factor of the traveling standard reported by th e participants is well within their reported uncertainties (k = 2), which a re between 0.5% and 3.1%.