D. Janik et al., Measurement techniques and results of an intercomparison for RF power in a3.5 mm coaxial line up to 26 GHz, IEEE INSTR, 48(2), 1999, pp. 166-168
Citations number
4
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
The methods and results of an international comparison are presented for me
asurements of RF power in a 3.5 mm coaxial line at seven frequencies up to
26 GHz. Three national laboratories of metrology having different measureme
nt techniques participated in the comparison. The agreement between the mea
n values of the calibration factor of the traveling standard reported by th
e participants is well within their reported uncertainties (k = 2), which a
re between 0.5% and 3.1%.