International comparison of thermal noise-temperature measurements at 2, 4, and 12 GHz

Citation
J. Randa et al., International comparison of thermal noise-temperature measurements at 2, 4, and 12 GHz, IEEE INSTR, 48(2), 1999, pp. 174-177
Citations number
10
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
48
Issue
2
Year of publication
1999
Pages
174 - 177
Database
ISI
SICI code
0018-9456(199904)48:2<174:ICOTNM>2.0.ZU;2-H
Abstract
We discuss an international comparison of thermal noise-power measurements (GTRF-92-2), which has recently been completed under the auspices of the Co nsultative Committee for Electricity and Magnetism (CCEM). The noise temper atures of two solid-state sources with GPC-7 connectors were measured at 2, 4, and 12 GHz at the national laboratories in France, Germany, the United K ingdom, and the United States. Good agreement was found among the results f rom the different laboratories, with all results agreeing within the expand ed uncertainties, which ranged from approximately 0.5% to 2.9%. The compari son was performed in accordance with the guidelines recently adopted by the Bureau International des Poids et Mesures (BIPM).