Extension of voltage range for power and energy calibrations

Citation
O. Petersons et al., Extension of voltage range for power and energy calibrations, IEEE INSTR, 48(2), 1999, pp. 439-443
Citations number
3
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
48
Issue
2
Year of publication
1999
Pages
439 - 443
Database
ISI
SICI code
0018-9456(199904)48:2<439:EOVRFP>2.0.ZU;2-O
Abstract
A special purpose ac voltage divider system with ratios of 600, 480, 360, 2 77, and 240 V to 120 V has been developed to extend the range of primary el ectric power calibrations from 120-600 V at 50 and 60 Hz, The developmental goal has been to realize ac voltage scaling within 5 mu V/V overall relati ve uncertainty. By employing a technique consisting of a special two-stage resistive divider compensated with an active circuit, the overall uncertain ty achieved after calibration is 4 mu V/V.