M. Mizugaki et al., Determination of leukotriene E-4 in human urine using liquid chromatography-tandem mass spectrometry, J CHROMAT B, 729(1-2), 1999, pp. 279-285
A Liquid chromatographic-tandem mass spectrometric (LC-MS-MS) method was de
veloped for the quantitation of urinary leukotriene E-4 (LTE4). LTE4 and it
s internal standard were extracted by solid-phase extraction and analysed u
sing LC-MS-MS in the selected reaction monitoring (SRM) mode. A good linear
response over the range of 10 pg. to 10 ng was demonstrated. The accuracy
of added LTE4 ranged from 97.0% to 108.0% with a mean and SD of 100.6+/-2.4
%. We detected LTE4 (63.1+/-18.7 pg/mg creatinine, n=10) in healthy human u
rine. This method can be used to determine LTE4 in biological samples. (C)
1999 Elsevier Science B.V. All rights reserved.