Color Doppler Flow Imaging (CDFI) of the vertebral arteries - The normal appearance, normal values and the proposal for the standards

Citation
A. Lovrencic-huzjan et al., Color Doppler Flow Imaging (CDFI) of the vertebral arteries - The normal appearance, normal values and the proposal for the standards, COLL ANTROP, 23(1), 1999, pp. 175-181
Citations number
17
Categorie Soggetti
Sociology & Antropology
Journal title
COLLEGIUM ANTROPOLOGICUM
ISSN journal
03506134 → ACNP
Volume
23
Issue
1
Year of publication
1999
Pages
175 - 181
Database
ISI
SICI code
0350-6134(199906)23:1<175:CDFI(O>2.0.ZU;2-5
Abstract
In order to assess mean. diameters and blood flow velocities (BFV), Color D oppler now Imaging (CDFI) of vertebral arteries (VA) was performed. Five hu ndred and ninety six persons without carotid disease or symptoms related to vertebrobasilar system were analyzed by CDFI of VA. Mean right VA diameter was 3.37+/-0.6 mm and left 3.55+/-0.61 mm. Women had thinner VA (p < 0.05) . Left VA was wider (p < 0.05). Mean. right BFV was 48.31+/-14.09 cm/s and left 48.93+/-13.94 cm/s. Females had higher BFV (p < 0.05). BFV didn't very with age (p > 0.05). The VA hypoplasia was present in 2.34%, asymmetry in 15% (left VA dominant in 64%). Visualisation of V1 and V2 segment was possi ble in 100% and of the origin in 81.7% on the right, and 80.7% on the left side. CDFI is a reliable method for evaluation of VA. Left VA was wider. Wo men had thinner VA. Hypoplasia was present in 2.34% and asymmetry in 15%.