Ne. Lee et al., Measured water transmission curves and calculated zero field size tumor maximum ratios for 4, 6, and 15 MV x-rays, MED PHYS, 26(6), 1999, pp. 956-961
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
A multihole diverging Cerrobend plug for megavoltage energies was used to m
easure water transmission values at different locations in a 20x20 cm field
at 100 cm source-to-axis distance (SAD) for 4, 6, and 15 MV therapy photon
beams. The transmission curves in water were measured at 25 locations acro
ss the 20x20 held, and each location was separated by 5 cm at the isocenter
. Each transmission value was made using a 0.3175 cm diameter (0.079 cm(2)
area) hole of 20 cm length at the central axis (CAX). The small field measu
red transmission curve in water was used to derive the zero field size tumo
r maximum ratio (TMR) and the primary photon exposure spectrum as a functio
n of energy at depth. The exposure spectrum was used to find an effective p
hoton energy and linear attenuation coefficient at depth and at different l
ocations in the field. These values were found to vary with location in the
field. (C) 1999 American Association of Physicists in Medicine.