Measured water transmission curves and calculated zero field size tumor maximum ratios for 4, 6, and 15 MV x-rays

Citation
Ne. Lee et al., Measured water transmission curves and calculated zero field size tumor maximum ratios for 4, 6, and 15 MV x-rays, MED PHYS, 26(6), 1999, pp. 956-961
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
26
Issue
6
Year of publication
1999
Pages
956 - 961
Database
ISI
SICI code
0094-2405(199906)26:6<956:MWTCAC>2.0.ZU;2-3
Abstract
A multihole diverging Cerrobend plug for megavoltage energies was used to m easure water transmission values at different locations in a 20x20 cm field at 100 cm source-to-axis distance (SAD) for 4, 6, and 15 MV therapy photon beams. The transmission curves in water were measured at 25 locations acro ss the 20x20 held, and each location was separated by 5 cm at the isocenter . Each transmission value was made using a 0.3175 cm diameter (0.079 cm(2) area) hole of 20 cm length at the central axis (CAX). The small field measu red transmission curve in water was used to derive the zero field size tumo r maximum ratio (TMR) and the primary photon exposure spectrum as a functio n of energy at depth. The exposure spectrum was used to find an effective p hoton energy and linear attenuation coefficient at depth and at different l ocations in the field. These values were found to vary with location in the field. (C) 1999 American Association of Physicists in Medicine.