Limitation of peak fitting and peak shape methods for determination of activation energy of thermoluminescence glow peaks

Citation
Cm. Sunta et al., Limitation of peak fitting and peak shape methods for determination of activation energy of thermoluminescence glow peaks, RADIAT MEAS, 30(2), 1999, pp. 197-201
Citations number
3
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
RADIATION MEASUREMENTS
ISSN journal
13504487 → ACNP
Volume
30
Issue
2
Year of publication
1999
Pages
197 - 201
Database
ISI
SICI code
1350-4487(199904)30:2<197:LOPFAP>2.0.ZU;2-3
Abstract
This paper shows the limitation of general order peak fitting and peak shap e methods for determining the activation energy of the thermoluminescence g low peaks in the cases in which retrapping probability is much higher than the recombination probability and the traps are filled up to near saturatio n level. Right values can be obtained when the trap occupancy is reduced by using small doses or by post-irradiation partial bleaching. This limitatio n in the application of these methods has not been indicated earlier. In vi ew of the unknown nature of kinetics in the experimental samples, it is rec ommended that these methods of activation energy determination should be ap plied only at doses well below the saturation dose. (C) 1999 Elsevier Scien ce Ltd. All rights reserved.