Correction of CAPD catheter displacement using gastric biopsy forceps: thepush-pull method

Citation
N. Ohira et al., Correction of CAPD catheter displacement using gastric biopsy forceps: thepush-pull method, INT J ARTIF, 22(4), 1999, pp. 202-204
Citations number
5
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology
Journal title
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS
ISSN journal
03913988 → ACNP
Volume
22
Issue
4
Year of publication
1999
Pages
202 - 204
Database
ISI
SICI code
0391-3988(199904)22:4<202:COCCDU>2.0.ZU;2-0
Abstract
We corrected malpositioned continuous ambulatory peritoneal dialysis cathet ers in six patients using a new technique named the "push-pull method" A ga stric biopsy forceps was advanced through the catheter to near ifs tip. Aft er manipulating the tip of the forceps through the abdominal wall, the forc eps was opened and pulled out slowly Repeated insertion and removal of the forceps induced the catheter to return to the pelvic cavity. This push-pull method was successful for Swan neck straight (n=2) and coiled (n=4) cathet ers in all patients. The time required for the procedure was only 5-10 minu tes and there were no complications.