Hip prothesis: an in vitro wear protocol based on a comparison between gravimetric and profilometric analysis

Citation
S. Affatato et al., Hip prothesis: an in vitro wear protocol based on a comparison between gravimetric and profilometric analysis, INT J ARTIF, 22(4), 1999, pp. 226-229
Citations number
16
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology
Journal title
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS
ISSN journal
03913988 → ACNP
Volume
22
Issue
4
Year of publication
1999
Pages
226 - 229
Database
ISI
SICI code
0391-3988(199904)22:4<226:HPAIVW>2.0.ZU;2-G
Abstract
Ultra high molecular weight polyethylene acetabular cups were analysed by m eans of a shadowgraph method (using a profile projector) to measure linear wear: The results were compared with those of previous wear tests performed on a hip joint simulator Twelve polyethylene acetabular cups were analysed . The specimens were evaluated visually for evidence of polyethylene wear. Ex amination of the polyethylene inner surface did not reveal evidence of surf ace failure such as delamination, fatigue cracks or scratches. Volumetric wear was calculated using a formula based on dimensional change due to the penetration of the femoral head in the acetabular cup. it was fo und to be of the same order of magnitude as the wear obtained in in vitro e xperimental tests.