E. Kongsgaard et al., TEMPERATURE GUIDED RADIOFREQUENCY CATHETER ABLATION OF MYOCARDIUM - COMPARISON OF CATHETER TIP AND TISSUE TEMPERATURES IN-VITRO, PACE, 20(5), 1997, pp. 1252-1260
Temperature monitoring during RF ablation has been proposed as a means
of controlling the creation of the lesion. However, in vivo studies h
ave shown poor correlation between lesion size and catheter Sip temper
ature. Thus, we hypothesized a difference between catheter tip and tis
sue temperatures during RF catheter ablation, and that this difference
may depend on flow passing the ablation site, tip electrode length, a
nd catheter-tissue orientation. In vitro studies were performed using
four different ablation catheters (tip electrode length: 2, 4, or 6 mm
) with a thermistor or a thermocouple as temperature sensor. Set tempe
rature was 70 degrees C and pulse duration was 30 seconds. Pieces of p
orcine left ventricle were immersed in a bath of isotonic saline-dextr
ose solution at 37 degrees C. The ablation catheters were positioned p
erpendicularly, obliquely, or parallel to the endocardium. A temperatu
re sensor was inserted from the epicardial side and positioned 1 mm be
neath the catheter-tissue interface. Experiments were made with a flow
of 200 mL/min passing the ablation site or with no flow. The catheter
tip and tissue temperatures differed significantly (P < 0.0001) durin
g ablation. This difference increased with time, with flow passing the
ablation site, with the length of the tip electrode, and when the cat
heter was positioned perpendicularly or obliquely to the endocardium a
s compared to the parallel catheter-tissue orientation (P < 0.05). In
conclusion, the tissue temperature may far exceed the catheter tip tem
perature, and intramyocardial superheating resulting in steam formatio
n and popping may occur despite a relatively low catheter tip temperat
ure.