J. Ivask et M. Kaljurand, Reduction of the detection limits of anions in polar ice core analysis using correlation ion chromatography with detector signal processing, J CHROMAT A, 844(1-2), 1999, pp. 419-423
Non-suppressed ion chromatography is usually a sensitivity deficit for pola
r ice core anionic content analysis. In the present study a method for sign
al improvement was tested. Ice core samples were injected into the chromato
graph according to a pseudo-random binary sequence. The chromatogram was ca
lculated by correlating the detector signal with the input sequence. The re
sults demonstrated that the method used gives increased sensitivity and is
applicable for performing ice core analysis but also requires high reproduc
ibility of injection system, preliminary sample modification and relatively
large sample volume. (C) 1999 Elsevier Science B.V. All rights reserved.