Radiometry using synchrotron radiation at PTB

Authors
Citation
M. Richter et G. Ulm, Radiometry using synchrotron radiation at PTB, J ELEC SPEC, 103, 1999, pp. 1013-1018
Citations number
30
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA
ISSN journal
03682048 → ACNP
Volume
103
Year of publication
1999
Pages
1013 - 1018
Database
ISI
SICI code
0368-2048(199906)103:<1013:RUSRAP>2.0.ZU;2-Z
Abstract
The facilities for radiometry using synchrotron radiation at the Physikalis ch-Technische Bundesanstalt are presented and the methods for absolute cali bration of radiation sources and detectors are described. Most of the measu rements are based on the use of two different types of primary standards, s torage rings as primary radiation source standards and cryogenic electrical substitution radiometers as primary detector standards. The facilities all ow us to offer radiometric calibrations in the spectral range from the UV t o x-rays with relative uncertainties down to 0.4%. (C) 1999 Elsevier Scienc e B.V. All rights reserved.