Evaluation of lot-to-lot consistency in ion exchange chromatography

Citation
Cb. Mazza et Sm. Cramer, Evaluation of lot-to-lot consistency in ion exchange chromatography, J LIQ CHR R, 22(11), 1999, pp. 1733-1758
Citations number
17
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES
ISSN journal
10826076 → ACNP
Volume
22
Issue
11
Year of publication
1999
Pages
1733 - 1758
Database
ISI
SICI code
1082-6076(1999)22:11<1733:EOLCII>2.0.ZU;2-N
Abstract
The reproducibility of ion exchange columns was investigated by rigorously evaluating critical chromatographic parameters as well as the performance o f these ion exchange columns under various modes of operation. A detailed s et of experiments were carried out to determine equilibrium Steric Mass Act ion parameters, HETP, column dead volume, total ion capacity, and porosity. In addition, the performance of these columns was examined under isocratic , analytical linear gradient, overloaded linear gradient, frontal, and disp lacement chromatographic conditions. While differences between the two lots were observed with isocratic experiments, the columns exhibited similar be havior under gradient, frontal, and displacement conditions. Furthermore, s imulations obtained with the SMA model were shown to predict column perform ance under linear and non-linear conditions.